Odom MJ, Lee YM, Zuckerman SL, Apple RP, Germanos T, Solomon GS, Sills AK. J. Surg. Orthop. Adv. 2016; 25(2): 93-98.

Affiliation

Vanderbilt University School of Medicine, Nashville, Tennessee.

Copyright

(Copyright © 2016, Data Trace Publishing)

DOI

unavailable

PMID

27518293

Abstract

This study evaluated the test-retest reliability of a novel computer-based, portable balance assessment tool, the Equilibrate System (ES), used to diagnose sports-related concussion. Twenty-seven students participated in ES testing consisting of three sessions over 4 weeks. The modified Balance Error Scoring System was performed. For each participant, test-retest reliability was established using the intraclass correlation coefficient (ICC). The ES test-retest reliability from baseline to week 2 produced an ICC value of 0.495 (95% CI, 0.123-0.745). Week 2 testing produced ICC values of 0.602 (95% CI, 0.279-0.803) and 0.610 (95% CI, 0.299-0.804), respectively. All other single measures test-retest reliability values produced poor ICC values. Same-day ES testing showed fair to good test-retest reliability while interweek measures displayed poor to fair test-retest reliability. Testing conditions should be controlled when using computerized balance assessment methods. ES testing should only be used as a part of a comprehensive assessment.