SAFETYLIT WEEKLY UPDATE

We compile citations and summaries of about 400 new articles every week.
RSS Feed

HELP: Tutorials | FAQ
CONTACT US: Contact info

Search Results

Journal Article

Citation

Di Pierro GB, Besmer I, Hefermehl LJ, Beatrice J, Danuser H, Beutler J, Mattei A. Eur. Urol. 2010; 58(4): 626-628.

Affiliation

Klinik für Urologie, Luzerner Kantonsspital, Lucerne, Switzerland.

Copyright

(Copyright © 2010, Elsevier Publishing)

DOI

10.1016/j.eururo.2010.07.030

PMID

20702029

Abstract

We report the first case of intra-abdominal combustion involving the plastic covering of monopolar scissors secondary to use of incorrect gas (oxygen O(2) instead of carbon dioxide CO(2)) during robot-assisted laparoscopic radical prostatectomy (RALP). The insufflating system was connected to a provisional O(2) gate into the operating theater. A patient underwent RALP and extended pelvic lymph node dissection for localized prostate cancer, according to standard technique. Approximately 1.5h after the start of surgery, flames arose from the scissor tips during monopolar coagulation. After extinguishing the fire, we promptly withdrew and changed instruments before recognizing and resolving the cause of the incident. The procedure was carried out without patient injury, and the postoperative period was uneventful.


Language: en