SAFETYLIT WEEKLY UPDATE

We compile citations and summaries of about 400 new articles every week.
RSS Feed

HELP: Tutorials | FAQ
CONTACT US: Contact info

Search Results

Journal Article

Citation

Hsiung MC, Chang YC, Wei J, Lan GY, Lee KC, Chang CY, Ou CH, Chou YP, Tsai SK. Echocardiography 2010; 27(5): 587-589.

Affiliation

Division of Cardiovascular Disease, Cheng-Hsin General Hospital, Taipei, Taiwan.

Copyright

(Copyright © 2010, John Wiley and Sons)

DOI

10.1111/j.1540-8175.2009.01126.x

PMID

20374269

Abstract

Endovascular stent grafts have become increasingly used in the management of arterial trauma, chronic vascular occlusive diseases, and deep vein thrombosis (DVT). Here, we describe a case of stent migration to the right heart due to a motor vehicle accident in a patient with iliac venous stent implanted for venous stenosis. Computed tomography (CT) scan imaging identification of a metallic stent was superior to the standard two-dimensional transesophageal echocardiography (TEE) techniques without imaging artifact from reverberation. However, real time three-dimensional (3D) TEE can precisely defined a metallic stent in right heart oriented with shape and direction without consuming time compared to that of CT scanning. The use of 3DTEE can facilitate identification of intracardiac metallic foreign bodies perioperatively in a patient with accident and emergent unstable condition.


Language: en

NEW SEARCH


All SafetyLit records are available for automatic download to Zotero & Mendeley
Print