TY - JOUR PY - 2021// TI - Selected ion flow tube-mass spectrometry (sift-ms) as an alternative to gas chromatography/mass spectrometry (gc/ms) for the analysis of cyclohexanone and cyclohexanol in plasma JO - ACS Omega A1 - Hastie, C. A1 - Thompson, A. A1 - Perkins, M. A1 - Langford, V.S. A1 - Eddleston, M. A1 - Homer, N.Z.M. SP - 32818 EP - 32822 VL - 6 IS - 48 N2 - Self-poisoning with professional agricultural pesticide products is responsible for about 20% of global suicide, with most cases occurring in South Asia and China. Treatment of severe poisoning involves long-term intensive clinical care and is often unsuccessful. Solvent co-formulants (such as cyclohexanone) also contribute to mortality themselves or via more toxic metabolic products (such as cyclohexanol). Faster detection of co-formulants could aid earlier identification of pesticide poisoning and faster intervention, reducing mortality. Conventional analysis of volatiles in blood uses headspace (HS)-GC/MS. This paper evaluates SIFT-MS, a direct MS technique that provides higher sample throughput than GC/MS, as a potential tool for cyclohexanone and cyclohexanol analysis in plasma. Both instruments were calibrated using a conventional approach prior to analysis of each porcine plasma sample on both instruments. Comparative data were evaluated using Bland-Altman plots, demonstrating that the techniques were in good agreement. Compared with GC/MS, SIFT-MS provides fourfold higher sample throughput and shows great promise as an alternative analytical tool. ©
Language: en
LA - en SN - 2470-1343 UR - http://dx.doi.org/10.1021/acsomega.1c03827 ID - ref1 ER -