TY - JOUR PY - 2020// TI - Development of a fast testing device for eye guards JO - China safety science journal (CSSJ) A1 - Shang, J. SP - 186 EP - 190 VL - 30 IS - 10 N2 - In order to improve quality control of eye guards and reduce testing period so as to guarantee safety of human eyes and promote export of products, luminous transmittance requirements and testing methods in domestic and foreign standards were analyzed. Then, based on broadband light source measurement method stipulated in the latest published international standard, a fast testing device was developed according to actual demand. Finally, it was calibrated by standard oculars. The results show that luminous transmittance requirements and testing methods vary significantly in standards of different countries. The newly developed testing device is accurate and fast, so it can be widely used as a standard testing instrument to measure luminous transmittance for eye guards. © 2020 China Safety Science Journal. All rights reserved.

Language: zh

LA - zh SN - 1003-3033 UR - http://dx.doi.org/10.16265/j.cnki.issn1003-3033.2020.10.026 ID - ref1 ER -