TY - JOUR PY - 2015// TI - A round robin test of piezoresistive shock accelerometers between shock and centrifuge calibrations JO - Transactions of Society of Automotive Engineers of Japan A1 - Nozato, Hideaki A1 - Oota, Akihiro A1 - Kokuyama, Wataru A1 - Chiba, Sakae A1 - Shimizu, Masao A1 - Kudo, Hiroyuki A1 - Maki, Kenji A1 - Tada, Ryuuichi A1 - Mochizuki, Tomohiro A1 - Iso, Shouji SP - 461 EP - 466 VL - 46 IS - 2 N2 - A round robin test of piezoresistive accelerometers was carried out to confirm a consistency between shock and centrifuge calibrations among five participants; AIST, KYOWA, TTDC, JARI and NICS. AIST evaluated the accelerometers using shock calibration system in compliance with ISO 16063-13, and the shock calibration included various frequency components. On the other hand, the others evaluated them using centrifuge calibration system in compliance with ISO 5347-7, and the centrifuge calibration based on DC (direct current) measurement. Finally, the shock and centrifuge calibrations were implemented from 1000 m/s2 to 5000 m/s2, and the consistency using Z scores with the standard deviation of 0.9 % was satisfied among five participants.

Language: ja

LA - ja SN - 0287-8321 UR - http://dx.doi.org/10.11351/jsaeronbun.46.461 ID - ref1 ER -