%T A randomised crossover comparison of manikin ventilation through Soft Seal®, i-gel™ and AuraOnce™ supraglottic airway devices by surf lifeguards
%J Anaesthesia
%D 2014
%A Adelborg, K.
%A Al-Mashhadi, R. H.
%A Nielsen, L. H.
%A Dalgas, C.
%A Mortensen, M. B.
%A Løfgren, B.
%V 69
%N 4
%P 343-347
%X Forty surf lifeguards attempted to ventilate a manikin through one out of three supraglottic airways inserted in random order: the Portex® Soft Seal®; the Intersurgical® i-gel™; and the Ambu® AuraOnce™. We recorded the time to ventilate and the proportion of inflations that were successful, without and then with concurrent chest compressions. The mean (SD) time to ventilate with the Soft Seal, i-gel and AuraOnce was 35.2 (7.2)s, 15.6 (3.3)s and 35.1 (8.5) s, respectively, p < 0.0001. Concurrent chest compression prolonged the time to ventilate by 5.0 (1.3-8.1)%, p = 0.0072. The rate of successful ventilations through the Soft Seal (100%) was more than through the AuraOnce (92%), p < 0.0001, neither of which was different from the i-gel (97%). The mean (SD) tidal volumes through the Soft Seal, i-gel and AuraOnce were 0.65 (0.14) l, 0.50 (0.16) l and 0.39 (0.19) l, respectively. Most lifeguards (85%) preferred the i-gel. Ventilation through supraglottic airway devices may be considered for resuscitation by surf lifeguards.<p /> <p>Language: en</p>