%0 Journal Article
%T An in vitro study of the accuracy and precision of Breathalyzer models 900, 900A, and 1000
%J Journal of forensic sciences
%D 1985
%A Caplan, Y. H.
%A Yohman, D. T.
%A Schaefer, J. A.
%V 30
%N 4
%P 1058-1063
%X Ninety Breathalyzer instruments (Model 1000) and twenty instruments (Models 900, 900A) were studied using a protocol described by the Department of Transportation's "Standard for Devices to Measure Breath Alcohol." Although the mean of each of three concentrations tested (0.05, 0.10, and 0.15 g/210 L) compared favorably in both series, the standard deviation was consistently higher for the Model 1000 instruments. The Model 1000 instruments also produced a significant number of test results which exceeded the normally expected scientific deviation.<p /> <p>Language: en</p>
%G en
%I John Wiley and Sons
%@ 0022-1198
%U http://dx.doi.org/