@article{ref1,
title="A round robin test of piezoresistive shock accelerometers between shock and centrifuge calibrations",
journal="Transactions of Society of Automotive Engineers of Japan",
year="2015",
author="Nozato, Hideaki and Oota, Akihiro and Kokuyama, Wataru and Chiba, Sakae and Shimizu, Masao and Kudo, Hiroyuki and Maki, Kenji and Tada, Ryuuichi and Mochizuki, Tomohiro and Iso, Shouji",
volume="46",
number="2",
pages="461-466",
abstract="A round robin test of piezoresistive accelerometers was carried out to confirm a consistency between shock andcentrifuge calibrations among five participants; AIST, KYOWA, TTDC, JARI and NICS. AIST evaluated theaccelerometers using shock calibration system in compliance with ISO 16063-13, and the shock calibration includedvarious frequency components. On the other hand, the others evaluated them using centrifuge calibration system incompliance with ISO 5347-7, and the centrifuge calibration based on DC (direct current) measurement. Finally, theshock and centrifuge calibrations were implemented from 1000 m/s2 to 5000 m/s2, and the consistency using Z scoreswith the standard deviation of 0.9 % was satisfied among five participants.<p /><p>Language: ja</p>",
language="ja",
issn="0287-8321",
doi="10.11351/jsaeronbun.46.461",
url="http://dx.doi.org/10.11351/jsaeronbun.46.461"
}