@article{ref1,
title="Quality assurance in breath-alcohol analysis",
journal="Journal of analytical toxicology",
year="1994",
author="Dubowski, Kurt M.",
volume="18",
number="6",
pages="306-311",
abstract="Evidential breath-alcohol testing requires an adequate quality assurance (QA) program to safeguard the testing process and validate its results. A comprehensive QA program covers (a) test subject preparation and participation; (b) the analysis process; (c) test result reporting and records; (d) proficiency testing, inspections, and evaluations; and (e) facilities and personnel aspects. Particularly important are the following necessary scientific safeguards as components of quality control: (a) a pretest deprivation-observation period of at least 15 minutes; (b) blank tests immediately preceding each breath-collection step; (c) analysis of at least duplicate breath specimens; and (d) a control test accompanying every subject test. These safeguards have withstood adversarial challenges in the judicial system for more than 30 years.<p /><p>Language: en</p>",
language="en",
issn="0146-4760",
doi="",
url="http://dx.doi.org/"
}