@article{ref1,
title="Alternating current electrocution detection and termination by an implantable cardioverter defibrillator",
journal="Pacing and clinical electrophysiology",
year="1997",
author="Mehdirad, A. and Love, C. and Nelson, S. and Schaal, S. and Collins, J. and Huffman, K.",
volume="20",
number="7",
pages="1885-1886",
abstract="A patient with an ICD accidentally grasped a power line and was electrocuted. He was unable to release the cable during electrocution though he remained conscious. After receiving a shock from his ICD, the powerline was released. ICD interrogation revealed inappropriate detection of alternating current and delivery of a shock.<p /><p>Language: en</p>",
language="en",
issn="0147-8389",
doi="",
url="http://dx.doi.org/"
}