@article{ref1,
title="The assessment of plutonium and americium in contaminated wounds with high energy resolution semiconductor detectors",
journal="Applied radiation and isotopes",
year="1995",
author="Genicot, J. L. and Hardeman, F. and Oberstedt, S.",
volume="46",
number="3",
pages="199-203",
abstract="A simple to use method was installed for the measurement of wounds contaminated with plutonium and americium by using a semiconductor detector. This technique does not explicitly require the knowledge of the detector efficiency but uses a plant 241Am calibration source. A computer programme has also been developed for the quantification of the contamination according to the shape and the depth of the contaminant in the wound.<p /><p>Language: en</p>",
language="en",
issn="0969-8043",
doi="",
url="http://dx.doi.org/"
}