@article{ref1,
title="Radiated electric field measurements in U.S. Army helicopters",
journal="Aviation, space, and environmental medicine",
year="1992",
author="Bruckart, J. E.",
volume="63",
number="11",
pages="1019-1023",
abstract="Aircraft systems and medical devices generate electromagnetic fields. Electromagnetic interference (EMI) can cause faulty operation of aircraft systems or medical devices and endanger patients or aircraft crewmembers. A ground and inflight study was conducted to describe the electromagnetic fields in typical operations. Broadband isotropic field sensors measured electric fields from 5 kHz to 3 MHz, 3 to 500 MHz, and 0.5 MHz to 6 GHz. Fields were measured at 0.5 m space intervals in JOH-58A, JUH-1H, and JUH-60A helicopters with systems off, operating RPM, 5-ft hover, 50-ft hover, and cruise. Electric fields in the environment were homogeneous and less than 0.1 V/m. Fields in the helicopters increased during ascent, but remained less than 2 V/m except during radio transmissions. EMI effect of the Physio Control Lifepack 8 was demonstrated during FM radio transmission. The results are useful in evaluating electromagnetic emissions and predicting operations that may result in an inflight malfunction of a medical device or aircraft system.<p /> <p>Language: en</p>",
language="en",
issn="0095-6562",
doi="",
url="http://dx.doi.org/"
}