@article{ref1,
title="An in vitro study of the accuracy and precision of Breathalyzer models 900, 900A, and 1000",
journal="Journal of forensic sciences",
year="1985",
author="Caplan, Y. H. and Yohman, D. T. and Schaefer, J. A.",
volume="30",
number="4",
pages="1058-1063",
abstract="Ninety Breathalyzer instruments (Model 1000) and twenty instruments (Models 900, 900A) were studied using a protocol described by the Department of Transportation's &quot;Standard for Devices to Measure Breath Alcohol.&quot; Although the mean of each of three concentrations tested (0.05, 0.10, and 0.15 g/210 L) compared favorably in both series, the standard deviation was consistently higher for the Model 1000 instruments. The Model 1000 instruments also produced a significant number of test results which exceeded the normally expected scientific deviation.<p /> <p>Language: en</p>",
language="en",
issn="0022-1198",
doi="",
url="http://dx.doi.org/"
}