@article{ref1,
title="Response of breath-alcohol analyzers to acetone",
journal="Journal of analytical toxicology",
year="1983",
author="Dubowski, Kurt M. and Essary, Natalie A.",
volume="7",
number="5",
pages="231-234",
abstract="Quantitative evidential breath-alcohol analyzers and screening testers in each of the five current categories of analytical principle were examined for response to dynamically-generated acetone vapor concentrations of 3, 100, 150, 350, and 600 micrograms/210 Liters. Nine of the 13 instruments tested were unaffected by acetone at any of these concentrations; three solid-state (Taguchi) sensing devices and one device employing single-wavelength infrared spectrometry displayed responses to acetone at the two highest tested concentrations. In view of a breath-acetone literature survey for ambulatory subjects and other considerations, response of the tested devices to acetone is not considered to be a significant problem in breath-alcohol analysis for traffic law enforcement purposes.<p /><p>Language: en</p>",
language="en",
issn="0146-4760",
doi="",
url="http://dx.doi.org/"
}